Quick start guide:
- Spray the chamber and glass slide holder with ethanol (70%)
- Rinse the chamber and glass slide holder well with distilled water
- Leave the chamber to dry in a sterile place (until the experiment starts)
- Take the glass slide holder into the fume hood for cell seeding.
Cell seeding:
- Place the glass slide holder in a suitably sized petri dish / similar
- If there is distilled water left in the glass slide holder wells, use a pipette to remove the remaining water.
- Seed 600uL at 0.67 * 10^5 cells/ml
- Put the lid on the petri dish and place it in an incubator overnight.
Here is a diagram of the experimental setup
General cleaning protocol
Equipment
- Vitro.alive (Vitroscope)
- 1x adapter
- 1x chamber
- 2x locking pin
- 6x silicone O-ring (already installed)
- 1x silicone filling tube
- 1x power cable
- 1x peristaltic pump (within range 0.015- 7.5ml/min)
- 3x silicone tubes (compatible with the pump)
- 1x inverted microscope
- 1x laminar flow hood
- 1x petri dish (150 mm) (Sarstedt) / plastic box